Continuous video recording with simultaneous amplitude-integrated EEG monitoring to improve seizure recognition in newborns

Alternative Title

Abstract

INTRODUCTION: Seizures in newborn infants are considered a neurological emergency requiring prompt treatment to limit exacerbation of brain injury. Digital monitors incorporating limited channel raw EEG and amplitude-integrated EEG (aEEG/EEG) are widely used. This study aimed to determine if continuous video recording with simultaneous aEEG/EEG recording enhances seizure recognition. METHODS: Newborns at risk of seizures who underwent neuromonitoring with aEEG/EEG were prospectively recruited to an observational study in a tertiary neonatal centre. Video recordings were commenced after obtaining written consent from parents. RESULTS: Simultaneous video recordings with aEEG/EEG were obtained in 15/47 newborns recruited to study. A total of 116 electrographic seizure episodes were detected on aEEG/EEG when a total of 56 episodes of abnormal movements were noted on video recordings. Only 8 of these abnormal movements had simultaneous electrographic seizures on aEEG. CONCLUSION: Use of simultaneous video and aEEG/EEG recordings in newborns at risk of seizures is feasible. It not only assists confirmation of the presence of seizures but may also help in identifying movements associated with abnormal neurology that are not seizures.

Description

Citation

Publisher

License

Journal

Frontiers in Pediatrics

Volume

13

Issue

PubMed ID

DOI

ISSN

EISSN